Assessment of the dynamic range in dynamic contrast-enhanced magnetic resonance imaging breast examinations

Introduction

Accurate dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) protocol evaluation is necessary to ensure reliable classification of contrast-agent (CA) uptake curves. This work presents a novel retrospective method to assess the dynamic range that constructs CA uptake curves using enhancement of the internal mammary artery.

Methods

Routine clinical breast examinations were performed using 3D fat-suppressed spoiled gradient-echo sequences (1.5T). Retrospective analysis was approved by the Clinical Audit Committee. Five different protocols were evaluated (10°, 14° and 18° flip angles (FAs), radial or linear k-space sampling), with seven to 10 patients in each group (n = 45). CA uptake curves were constructed from a standardised axial slice through the right mammary artery, and maximum relative enhancement (E max ) and time-to-peak enhancement (T max ) were measured for each examination [1]. E max and T max were compared between protocols (ANOVA/Mann-Whitney; P < 0.05 indicating significance). For each protocol, calculated values of maximum relative enhancement (E calc ) were derived from the Bloch equations and compared with E max to validate the results.

Results

A lower FA and radial k-space sampling resulted in a statistically significant decrease in E max (P < 0.0001 and P = 0.001, respectively). Radial protocols exhibited greater T max than linear protocols at FAs of both 14° (P = 0.025) and 18° (P < 0.0001). E calc was found to increase with FA as expected, with good agreement between E calc and E max .

Conclusion

Significant differences were found between patient groups with only small alterations in protocol. Observations agreed with expected results, validating this method of retrospective analysis. The dynamic range is optimised at higher flip angles and with linear k-space sampling.

References

  1. Kuhl , et al: Radiology. 1999, 211: 101-110.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Cite this article

Ledger, A., Borri, M., Schmidt, M. et al. Assessment of the dynamic range in dynamic contrast-enhanced magnetic resonance imaging breast examinations. Breast Cancer Res 14 (Suppl 1), P39 (2012). https://doi.org/10.1186/bcr3294

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/bcr3294

Keywords