Accuracy of GE digital breast tomosynthesis versus supplementary mammographic views for diagnosis of screen-detected soft tissue breast lesions

  • Eleanor Cornford1,
  • Anne Turnbull2,
  • Jonathan James1,
  • Rachel Tsang1,
  • Tayeba Akram2,
  • Helen Burrell1,
  • Lisa Hamilton1,
  • Sarah Tennant1,
  • Mark Bagnall2,
  • Shama Puri2,
  • Graham Balls3,
  • Yan Chen4 and
  • Vivienne Jones5
Breast Cancer Research201517(Suppl 1):O1

https://doi.org/10.1186/bcr3757

Published: 5 November 2015

Introduction

The aim was to compare the accuracy of standard supplementary views and GE digital breast tomosynthesis (DBT) for assessment of soft tissue mammographic abnormalities.

Methods

Women recalled for further assessment of soft tissue abnormalities were recruited and received standard supplementary views (typically spot compression views) and two-view GE DBT. The added value of DBT in the assessment process was determined by analysing data collected prospectively by unblinded radiologists working up the cases.

Following anonymisation of cases, there was also a retrospective multireader review. The readers first read bilateral standard two-view digital mammography (DM) together with the supplementary mammographic views and gave a combined score for suspicion of malignancy on a five-point scale. The same readers then read bilateral standard two-view DM together with two-view DBT. Pathology data were obtained. Differences were assessed using ROC analysis.

Results

The study population was 342 lesions in 322 patients. Final diagnosis was malignant in 113 cases (33%) and benign/normal in 229 cases (67%). In the prospective analysis, the performance of two-view DM plus DBT was at least equivalent to the performance of two-view DM and standard mammographic supplementary views--area under the curve (AUC) was 0.946 and 0.922 respectively, which did not reach statistical significance. Similar results were obtained for the retrospective review--AUC was 0.900 (DBT) and 0.873 (supplementary views), which did not reach statistical significance.

Conclusion

The accuracy of GE DBT in the assessment of screen-detected soft tissue abnormalities is equivalent to the use of standard supplementary mammographic views.

Authors’ Affiliations

(1)
Nottingham Breast Institute
(2)
Royal Derby Hospital
(3)
Nottingham Trent University
(4)
Loughborough University
(5)
Northampton General Hospital

Copyright

© Cornford et al. 2015

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Advertisement