Table 3 Sensitivity of digital breast tomosynthesis and digital mammography in all quintiles

From: Impact of breast density on diagnostic accuracy in digital breast tomosynthesis versus digital mammography: results from a European screening trial

Quintile

Sensitivity DBT

% (n) (95% CI)

Sensitivity DM

% (n) (95% CI)

Difference DBT vs DM percentage points (95% CI)

p Value

PD 1

90.9% (20/22)

(72.2–97.5)

72.7% (16/22)

(51.8–86.8)

18.2

( − 4.6–41.0)

0.13

PD 2

82.1% (23/28)

(64.4–92.1)

64.3% (18/28)

(45.8–79.3)

17.8

( − 5.4–41.0)

0.13

PD 3

72.5% (29/40)

(57.2–83.9)

62.5% (25/40)

(47.0–75.8)

10.0

( − 10.5–30.5)

0.42

PD 4

83.9% (26/31)

(67.4–92.9)

64.5% (20/31)

(46.9–78.9)

19.4

( − 2.4–41.2)

0.11

PD 5

81.1% (30/37)

(65.8–90.5)

43.2% (16/37)

(28.7–59.1)

37.9

(15.8–60.0)

 < 0.001

Overall

81.0% (128/158)

(74.2–86.4)

60.1% (95/158)

(52.3–67.4)

20.9

(10.8–30.9)

 < 0.001

DA 1

85.7% (18/21)

(65.4–95.0)

81.0% (17/21)

(60.0–92.3)

4.8

( − 17.8–27.3)

 > 0.99

DA 2

82.1% (23/28)

(64.4–92.1)

75.0% (21/28)

(56.6–87.3)

7.1

( − 14.4–28.6)

0.69

DA 3

80.0% (24/30)

(62.7–90.5)

60.0% (18/30)

(42.3–75.4)

20.0

( − 3.2–43.2)

0.11

DA 4

76.7% (33/43)

(62.3–86.8)

51.2% (22/43)

(36.8–65.4)

25.6

(5.3–45.9)

0.003

DA 5

83.3% (30/36)

(68.1–92.1)

47.2% (17/36)

(32.0–63.0)

36.1

(14.1–58.1)

0.002

Overall

81.0% (128/158)

(74.2–86.4)

60.1% (95/158)

(52.3–67.4)

20.9

(10.8–30.9)

 < 0.001