Assessment of patient doses during mammography practice at Kenyatta National Hospital

Introduction

Breast cancer constitutes 21% of all cancer cases and ranks third according to cancer type in Kenya. Mammography therefore remains a powerful radiographic imaging technique for detecting and managing breast cancer. However, the active and radiosensitive glandular tissue calls for the need for an effective quality assurance programme.

Methods

A questionnaire method was developed and used in recording the displayed patient dose, compressed breast thickness and exposure factors. The average glandular dose, device performance, and film quality grading were also carried out at the largest mammography facility in Kenya.

Results

There were 3,264 films from 1,252 women of between 25 and 90 years old. The AGD per film was 2.14 (range 0.27 to 9.43) mGy for the CC projection and 2.44 (range 0.20 to 10.12) mGy for the MLO projection. A total 17% of CC and 30% of MLO films recorded doses above the recommended 3 mGy diagnostic reference level.

Conclusion

In view of the diagnostic dose finding of this study, the variation of mammography imaging techniques revealed the need for standards and optimisation of mammography practice in Kenya.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Cite this article

Wambani, J., Korir, G., Shyanguya, M. et al. Assessment of patient doses during mammography practice at Kenyatta National Hospital. Breast Cancer Res 13 (Suppl 1), P44 (2011). https://doi.org/10.1186/bcr2996

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/bcr2996

Keywords