Can radiation dose in mammography be further reduced by increasing the image quality?

Introduction

Efforts to reduce radiation dose should be linked to and limited by image quality and the radiation dose should only be lowered to levels compatible with image quality for adequate diagnosis; this is optimisation of mammography. Despite this link between image quality and mean glandular dose (MGD), the MGD is proposed only with regard to compression level achieved or population percentile, not image quality required/achieved.

Methods

A study of symptomatic breast units geographically spread over Ireland collected image quality and radiation dose data. The quantitative and qualitative data were analysed using mathematical modelling and SPSS statistics including ANOVA. High image quality was matched to lowest achievable radiation doses; inadequate images were discarded from the dataset for recommendations of achievable MGD.

Results

MGDs received by perfect images are significantly lower than the radiation doses received by inadequate images. This is seen in both digital images and film-screen images and for both mammographic projections: digital craniocaudal, F(3,978) = 2.841, P = 0.37; digital mediolateral oblique, F(3,977) = 4.896, P = 0.002; analogue craniocaudal, F(3,785) = 7.993, P < 0.001; analogue mediolateral oblique, F(3,783) = 7.961, P < 0.001). The mean MGD in mGy required to produce a perfect image in each of the categories: digital craniocaudal, 1.23 mGy; digital mediolateral oblique, 1.28 mGy; analogue craniocaudal, 2.10 mGy; analogue mediolateral oblique, 2.25 mGy.

Conclusion

The publication of average glandular dose must be linked to image quality achieved and the percentage of inadequate images in the data collection must be explicit. The radiation doses can be lowered further in digital imaging by greater training of radiographers to consistently achieve perfect mammography images.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Cite this article

O'Leary, D., Rainford, L. Can radiation dose in mammography be further reduced by increasing the image quality?. Breast Cancer Res 13 (Suppl 1), P21 (2011). https://doi.org/10.1186/bcr2973

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/bcr2973

Keywords