Mammographic image unsharpness: a predictable phenomenon?

Introduction

In 2009, the South West of Scotland Breast Screening Unit (SWSBSU) had to repeat 236 examinations due to mammographic image unsharpness. Such recalls were undesirable because of radiation safety issues, anxiety caused and cost. The aim of this study was to model predictors of recall that could be used in clinical practise to help reduce the numbers of repeat examinations required due to blurring.

Methods

This retrospective study compared two sample groups (n = 118), randomly selected from a cohort of women who attended the SWSBSU in 2009 (n = 16,194). Tests of significance were used to compare a range of variables in each group and logistic regression was employed to produce four predictor models for recall. Statistical analysis was carried out by the authors using SPSS version 18.

Results

A 12 mm increase in compressed breast thickness (CBT), a 50 mAs increase in exposure or a 350 ms increase in exposure time will double the odds of recall due to unsharpness. Also, an increase of 2 kVp, when imaging a breast with a CBT of 60 mm, will reduce the odds of recall by approximately 39%.

Conclusion

CBT is a major determinant of recall due to unsharpness. Within the context of the reported increase in average CBT [1] and rising national obesity, reconsideration of kVp selection criteria may be necessary to minimise the incidence of recall due to unsharpness.

References

  1. Robinson M, Kotre C: Trends in compressed breast thickness and radiation dose in breast screening mammography. Br J Radiol. 2008, 81: 214-218. 10.1259/bjr/90916004.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Cite this article

Coltart, A., Blythe, C. & Hair, M. Mammographic image unsharpness: a predictable phenomenon?. Breast Cancer Res 13 (Suppl 1), P1 (2011). https://doi.org/10.1186/bcr2953

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/bcr2953

Keywords